DIODE ARRAY SPECTROPHOTOMETER HP 8452A


Steps to run the Diode Array Spectrophotometer:
1) Turn the UV/Vis Spectrophotometer switch on.
2) Turn the computer on.
3) Select Microsoft Windows 2000.
4) Start the Chemstation by clicking the icon UV-Vis online.
5) Go to Instrument and turn the light on.
6) Go to file—> New Method /Load (Previous Method)
7) Method
—Fixed Wavelength Task: The Fixed Wavelength task allows the calculation of results from up to six wavelengths with or without back ground correction.
· Use wavelengths —Up to six single wavelength
·
Background correction—none
single reference wavelength
subtract average over a range
three-point drop line
·
Data Type—Transmittance
Absorbance
Derivative 1-4
· Display Spectrum—Sets the upper and lower wavelength limits for spectral display
—Spectrum/Peak: This task determines the maximum and minimum of the data points(y-values) in the defined wavelength range of the spectrum.
· Data Type—Sets the Spectral Processing to the selected type
—Transmittance
—Absorbance
—Derivative 1-4
· Display Spectrum—set the upper and lower wavelength limits for spectral display.
· Number of peaks—0 to 19
· Number of Valleys—0 to 19
—Ratio/Equation: This task calculates the result based upon a specified equation. The equation can contain absorbances from up to six wavelengths plus sample weight and volume information. The parameters are as follows:
· Use Wavelengths: Set it to the selected values. Up to six single wavelengths.
· Data Type: Set it to the selected values.
—Transmittance
—Absorbance
—Derivative 1-4
· Equation Name: Set the equation name-Numeric text
· Equation: Set the equation parameter-Numeric text
You can use the following mathematical operators for the equation:
+,-,/,*,ABS(exp),Ceil(exp),EXP(exp),FlOOR(exp),LN(exp)
LOG (exp), SQR (exp), SQRT (exp).
· Unit: Set the units for the result-Numeric text
· Use sample weight and volume: Switches on the weight and volume entry fields in the Sample Information dialog box.
Checked or unchecked.
· Prompt for sample information: Switches on the display of the Sample Information dialog box after sample measurement. Checked or unchecked.
· Display Spectrum: Sets the upper and lower wavelength limits for spectral display.
—Quantification: This task calculates calibration coefficients from the measured data of a set of standards, and then uses the calibration coefficients to determine the concentration of unknown sample.
· Use wavelengths —Single wavelength value.
·
Background correction—none
single reference wavelength
subtract average over range
three-point drop line
· Analyte name: Set the analyte name-numeric text.
· Calibration Curve Type: Set the calibration curve type.
—linear (Beer’s law)
—linear with offset
—quadratic
—quadratic with offset
· Concentration: set the unit of concentration-numeric text.
· Weight and Volume unit: Set the units of weight and volume; Switches on the weight and volume entry fields in the Sample Information dialog box-numeric text.
· Prompt for standard information: Switches on the display of the Standard Information dialog box after sample measurement. Checked or unchecked.
· Prompt for sample information: Switches on the display of the Sample Information dialog box after sample measurement. Checked or unchecked.
· Data Type: Sets the Spectral Processing to the selected type
—Transmittance
—Absorbance
—Derivative 1-4
· Display Spectrum: Sets the upper and lower wavelength limits for spectral display.
8) Go to Measure from the screen Menu.
· Blank (F4) to run the blank
· Automation Set up (F7)
9) Go to Instrument:
· Select sampling system—manula
—Sipper
—Multiple Cell
—Auto Sampler
press sample to run the sample
· Set up spectrophotometer—Wavelength range from 190-1100
—Integrated time-.5 sec
· Spectrum status
10) Math —Absorbance: It is the default method. Absorbance spectra is calculated using absorbance function.
—Transmittance: Transmittance spectra is calculated using transmittance function.
—Derivative: First, second, third and fourth order derivative spectra can be calculated. Derivative function is used.
—SPline
11) View—samples
—last blank spectrum
—math results
12) Mode—Standard
13) Config—Path: standard report
—data
—method
14) Help
15) After you are done go to instrument turn off the lamp, spectrophotometer and the computer.